Remote Control Commands
R&S
®
Scope Rider RTH
349User Manual 1326.1578.02 ─ 12
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Parameters:
<WidthY> Range: 0 to 8
Increment: 0.01
*RST: 0.1
Default unit: div
MASK:CHANnel<m>:CREatemask
Creates a mask from the envelope waveform of the selected waveform with the
defined width in x and y direction.
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Usage: Event
15.5.2
Mask Test
MASK:ONViolation[:SELection]....................................................................................... 349
MASK:RST....................................................................................................................349
MASK[:TESTstate]?........................................................................................................350
MASK:CHANnel<m>:RESult:FAIL:PERCentage?.............................................................. 350
MASK:CHANnel<m>:RESult:FAIL[:COUNt]?.....................................................................350
MASK:CHANnel<m>:RESult:PASS:PERCentage?............................................................ 350
MASK:CHANnel<m>:RESult:PASS[:COUNt]?................................................................... 351
MASK:CHANnel<m>:RESult:TOTL[:COUNt]?................................................................... 351
MASK:ELAPsedtime:TOTal?........................................................................................... 351
MASK:ELAPsedtime[:SECS]?......................................................................................... 352
MASK:ONViolation[:SELection] <SelectedActions>
Defines the action to be executed if a violation occurs.
Parameters:
<SelectedActions> NONE | STOP | BEEP | BPSTop
*RST: NONE
MASK:RST
Sets the counters of passed and failed acquisitions to Zero.
Usage:
Event
Mask Testing