Remote Control Commands
R&S
®
Scope Rider RTH
386User Manual 1326.1578.02 ─ 12
Parameters:
<Unit> C | F | K
METer:SENSe:TEMPerature:TRANsductor:RTD:TYPE <Unit>
Sets the adapter type for temperature measurements.
Parameters:
<Unit> PT100 | PT500
15.7.3 Relative Measurements
To perform relative measurements, set the relevant
METer:SENSe:<function>:NULL:STATe command to ON. By default, the reference
value is 0. Alternatively, you can use METer<m>:SENSe:RELative:STATe.
To change the reference value, use the METer:SENSe:<function>:NULL:VALue
command.
METer:SENSe:CAPacitance:NULL:STATe........................................................................ 386
METer:SENSe:CURRent:AC:NULL:STATe........................................................................386
METer:SENSe:CURRent:DC:NULL:STATe........................................................................386
METer:SENSe:RESistance:NULL:STATe.......................................................................... 386
METer:SENSe:TEMPerature:NULL:STATe........................................................................386
METer:SENSe:VOLTage:AC:NULL:STATe........................................................................ 386
METer:SENSe:VOLTage:DC:NULL:STATe........................................................................ 386
METer<m>:SENSe:NULL:STATe......................................................................................387
METer<m>:SENSe:RELative:STATe.................................................................................387
METer:SENSe:CAPacitance:NULL:VALue........................................................................ 387
METer:SENSe:CURRent:AC:NULL:VALue........................................................................387
METer:SENSe:CURRent:DC:NULL:VALue....................................................................... 387
METer:SENSe:VOLTage:AC:NULL:VALue........................................................................ 387
METer:SENSe:VOLTage:DC:NULL:VALue........................................................................ 387
METer:SENSe:RESistance:NULL:VALue..........................................................................388
METer:SENSe:TEMPerature:NULL:VALue........................................................................388
METer<m>:SENSe:NULL:VALU.......................................................................................388
METer:SENSe:CAPacitance:NULL:STATe <State>
METer:SENSe:CURRent:AC:NULL:STATe <State>
METer:SENSe:CURRent:DC:NULL:STATe <State>
METer:SENSe:RESistance:NULL:STATe <State>
METer:SENSe:TEMPerature:NULL:STATe <State>
METer:SENSe:VOLTage:AC:NULL:STATe <State>
METer:SENSe:VOLTage:DC:NULL:STATe <State>
Enables or disables the relative measurement. The reference value is defined using
the appropriate METer:SENSe:<function>:NULL:VALue command.