Remote Control Commands
R&S
®
Scope Rider RTH
409User Manual 1326.1578.02 ─ 12
COUNter<m>:CONFigure:FREQuency <Range>
This command sets all measurement parameters and trigger parameters to default val-
ues for frequency measurements.
Note: this command can be executed at any time, even if the counter is not yet activa-
ted. The parameters are stored internally and used when the counter is activated.
Suffix:
<m>
.
1..2
Parameters:
<Range> <numeric value> | MIN | MAX | DEF
Defines the measurement range of the input signal for the coun-
ter. If probes are connected, the MIN | MAX | DEF values are
adapted accordingly.
<numeric value>
Power range in V. If necessary, the next higher range is
selected.
MIN
100 mV
MAX
300 V
DEF
100 mV
Usage:
Setting only
COUNter<m>:CONFigure:VALue?
Returns the current configuration for the counter measurement defined by the most
recent COUNter<m>:CONFigure:FREQuency or COUNter<m>:MEASure:
FREQuency? command.
Suffix:
<m>
.
1..2
Return values:
<ConfigString> String containing the current measurement function (FREQ) and
the defined measurement range of the used input channel.
Example:
COUN1:CONF:VAL?
//Result: 'FREQ 3V'
Usage: Query only
COUNter<m>:FETCh?
Queries the currently measured value.
Suffix:
<m>
.
1..2
Counter Mode (R&S RTH-K33)