Remote Control Commands
R&S
®
Scope Rider RTH
447User Manual 1326.1578.02 ─ 12
Parameters:
<BitOrder> LSBF | MSBF
*RST: MSBF
BUS:UART:FRAMemode <FrameMode>
IDLE defines frames of several words in the data stream, which are defined by a time-
out between a stop bit and the next start bit. Enter the minimum timeout between two
frames using BUS:UART:TOUT.
Parameters:
<FrameMode> NONE | IDLE
*RST: NONE
BUS:UART:TOUT
<IdleTime>
Sets the minimum timeout between two frames if BUS:UART:FRAMemode is set to
IDLE.
Parameters:
<IdleTime> Range: 100E-9 to 1
Increment: 1E-3
*RST: 1E-3
Default unit: s
15.11.4.2 UART Trigger
TRIGger:UART:TYPE..................................................................................................... 447
TRIGger:UART:DATA..................................................................................................... 448
TRIGger:UART:DCONdition............................................................................................ 448
TRIGger:UART:TYPE <TriggerType>
Selects the trigger type for UART analysis.
Parameters:
<TriggerType> STBT | PCKS | DATA | PRER | STPerror | BRKC
STBT = start bit
PCKS = frame start
PRER = parity error
STPerror = stop error
BRKC = break condition
DATA:: data. To set up the trigger condition, use TRIGger:
UART:DATA, and TRIGger:UART:DCONdition.
For details, see "UART Trigger" on page 200.
*RST: STBT