Remote Control Commands
R&S
®
Scope Rider RTH
449User Manual 1326.1578.02 ─ 12
Suffix:
<m>
.
*
Selects the word.
Return values:
<State> OK | BREak | STERror | SPERror | PRERror | INSufficient
OK: the frame is valid.
BREak: break condition found. A start bit is not followed by a
stop bit, and the data line remains at logic 0 for longer than a
UART word.
STERror: start error, no start bit found.
SPERror: stop error, no stop condition found.
PRERror: parity error, which indicates a transmission error.
INSufficient: the frame is not completely contained in the
acquisition. The acquired part of the frame is valid.
*RST: OK
Usage: Query only
BUS:UART:WORD<m>:STARt?
Returns the start time of the specified symbol (word).
Suffix:
<m>
.
*
Selects the word.
Return values:
<StartTime> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS:UART:WORD<m>:STOP?
Returns the end time of the specified symbol (word).
Suffix:
<m>
.
*
Selects the word.
Return values:
<StopTime> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
Protocol Analysis