EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 468

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
468User Manual 1326.1578.02 ─ 12
Parameters:
<ChecksumError> ON | OFF
*RST: ON
TRIGger:LIN:DATA <Pattern>
Defines the data pattern as trigger condition. Enter the words in MSB first bit order.
Parameters:
<Pattern> String with max. 4 bytes in binary format. Characters 0, 1, and X
are allowed. If you define a pattern with incomplete byte, the
missing LSB are filled with X.
Example:
TRIG:LIN:TYPE IDDT
TRIG:LIN:IDEN 001001
TRIG:LIN:ICON EQU
TRIG:LIN:DCON EQU
TRIG:LIN:DATA '11111111000000001111'
TRIG:LIN:DATA?
<-- 11111111000000001111XXXX
Triggers when the id '001001' and the data
'11111111000000001111' is detected in the measured signal
TRIGger:LIN:DCONdition <Relation>
Defines how the specified data pattern is compared with the acquired signal.
Parameters:
<Relation> EQUal | NEQual
*RST: EQUal
TRIGger:LIN:ICONdition <Relation>
Defines how the specified identifier pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation> EQUal | NEQual
*RST: EQUal
TRIGger:LIN:IDENtifier <Pattern>
Specifies the identifier pattern to be found, in binary format. Enter the pattern in MSB
first bit order.
Parameters:
<Pattern> String with max. 7 characters. Characters 0, 1, and X are
allowed. If you define a pattern shorter than the pattern length,
the missing LSB are filled with X.