15.11.7.2 SENT Trigger
TRIGger:SENT:TYPE..................................................................................................... 478
TRIGger:SENT:DATA......................................................................................................478
TRIGger:SENT:DCONdition............................................................................................ 478
TRIGger:SENT:ICONdition..............................................................................................479
TRIGger:SENT:IDENtifier................................................................................................479
TRIGger:SENT:SCONdition.............................................................................................479
TRIGger:SENT:STATus...................................................................................................479
TRIGger:SENT:FCRCerror..............................................................................................479
TRIGger:SENT:IRFLength...............................................................................................480
TRIGger:SENT:PPERioderror..........................................................................................480
TRIGger:SENT:PULSeerror.............................................................................................480
TRIGger:SENT:SCRCerror..............................................................................................480
TRIGger:SENT:TYPE <TriggerType>
Sets the SENT trigger type.
Parameters:
<TriggerType> STOF | STAT | STDA | ID | IDDT | ERRC
STOF:
STAT: fast status
STDA: fast status and data
ID: slow Identifier
IDDT: slow Identifier and data
ERRC: error condition
*RST: STOF
TRIGger:SENT:DATA <Pattern>
Defines the data pattern as trigger condition. Enter the words in MSB first bit order.
Parameters:
<Pattern> String with max. 4 bytes in binary format. Characters 0, 1, and X
are allowed. If you define a pattern with incomplete byte, the
missing LSB are filled with X.
TRIGger:SENT:DCONdition <Relation>
Defines how the specified data pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation> UNUSed | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
*RST: EQUal