R&S SMB RF Level
1407.0806.32 4.33 E-2
Level - Menu
The Level menu is opened in the RF function block or using the MENU key under RF.
The offset-free level, level offset and level limit are set in the top section of the menu. The attenuator
mode is set in the Attenuator Settings section .The step width which is used when setting the level
using the rotary knob (with Variation Active On) is set in the User Variation section.
The offset-free level, attenuation mode, level offset and level limit are set in the top section of the menu.
Amplitude - RF Level
Sets the RF level of the RF output connector.
The level entered and displayed here corresponds to the level at the
RF output, i.e. any offset entry is not taken into consideration.
Remote-control command: n.a.
Note:
There is no GPIB command available for this setting. The GPIB
command SOUR:POW:LEV:IMM:AMPL sets the level of the Level
display, i.e. the level containing offset.