RF EMF R&S SMB
1407.0806.32 4.36 E-2
RF EMF
The power-on behavior of the R&S SMB Signal Generator and the level display in the display header
are set in the RF EMF menu.
EMF - Menu
The EMF menu is opened in the RF function block or using the MENU key under RF.
The power-on behavior of the R&S SMB Signal Generator and the level display in the display header
are set in the Power-On / EMF Settings section.
Power-On State - RF
Output
Selects the state which the RF output is to assume after the
instrument is switched on.
RF Off
The output is deactivated when the instrument is
switched on.
Remote-control command:
OUTP:PON OFF
Previous
Setting
When the instrument is switched on, the output
assumes the same state as it had when the
instrument was switched off.
Remote-control command:
OUTP:PON ON
Display Level as Voltage of
EMF - RF Level
Activates display of the signal level as voltage of the EMF (no-load
voltage). If this setting is deactivated, the level is displayed as a
voltage with 50 Ohm (presetting).
Remote-control command: n.a.