R&S SMB Alphabetical List of Device-Specific Error Messages
1407.0806.32 9.11 E-2
Alphabetical List of Device-Specific Error Messages
The following list contains all error messages specific of the instrument in alphabetical order. The
positive error codes mark the errors specific of the instrument.
The error text being entered into the error/event queue or being displayed is printed in bold face on the
left together with the error code. Below the error text, there is an explanation as to the respective error.
Note:
The index provides a list of the error messages sorted according to their error codes.
Adjustment data invalid (183)
Adjustment data are invalid and must be restored.
Remedy:
The adjustment data have to be generated again by an internal
or external adjustment or to be loaded into the instrument.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Adjustment data missing (182)
Adjustment data are missing.
Remedy:
The adjustment data have to be generated first by an internal or
external adjustment or to be loaded into the instrument.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Adjustment failed (180)
Adjustment could not be executed
Remedy:
The adjustment data have to be generated first by an internal or
external adjustment or to be loaded into the device (see section
Adjustment).
SCPI:
Device-specific error - sets bit 3 in the ESR register
Cannot access the EEPROM (202)
A error occurs when writing or reading a EEPROM.
Example:
The EEPROM is defect.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Cannot access hardware (200)
The data transmission to a module was unsuccessful.
Example:
The module is not installed, not properly installed or missing.
SCPI:
Device-specific error - sets bit 3 in the ESR register