General Instrument Settings  R&S SMB 
1407.0806.32 4.2  E-2 
General Instrument Settings 
Overview of General Instrument Settings 
The section "General Instrument Settings" describes the settings which do not directly affect signal 
generation. Most of these settings can only be accessed by means of menus which are opened using 
the front panel key.  
The general instrument settings affect various functions, such as storing instrument settings using the 
 
FILE 
 key or setting the IEC/IEEE-bus address in the menu of the  SETUP  key. The order in which the 
descriptions are given corresponds to the layout of the keys on the front panel of the R&S SMB (from 
top left to bottom right). 
 
Default Instrument Settings - Preset Key 
The  PRESET  key calls up a defined instrument setup. All parameters and switching states are preset 
(also those of inactive operating modes). The default instrument settings provide a reproducible initial 
basis for all other settings. 
Remote-control command: 
*RST 
When the instrument is switched on, it is not the preset state that is active, but rather the instrument 
state that was set before the instrument was switched off. An exeption is the state of the RF output. The 
state of the RF output power-on state can be set to be always off in the EMF menu of the RF block. 
User-defined instrument states can be stored and called up in the File menu. 
The following table gives an overview of the presets for the most important generator settings. The 
other presets can be found in the information accompanying the IEC/IEEE-bus commands (Chapter 
"Remote Control Commands - Description"). 
Table 4-1    Preset state - important generator settings 
Setting Preset value 
RF frequency   1 GHz 
RF level  RF output switched off 
Reference frequency  Internal, adjustment off 
Offsets 0 
Modulations Switched off 
Uninterrupted level setting  Switch off; level attenuator mode: AUTO 
Internal level control  Level ALC: AUTO 
User correction   Level Ucor: OFF 
LF output  Switched off 
Sweep Switched off 
IEC/IEEE-bus address  Not changed 
 
 
 
 