Example of Setup R&S SMB
1407.0812.62 2.8 E-2
Generation of an RF Frequency Sweep Signal
In the example, an RF frequency sweep is configured. Proceed as described below:
1. Activate default (preset) state.
2. Configure and activate RF frequency sweep. The settings are:
start frequency: 100 MHz
stop frequency: 500 MHz.
linear spacing with a step width of 1 MHz
dwell time 10 ms
single sweep mode
3. Activate RF signal.
4. Trigger single sweep
Step 1: Activate default (preset) state
Set a defined instrument state by pressing the
PRESET
key.
Step 2: Configure and activate RF frequency sweep
Select the RF block by turning the rotary knob.