R&S SMB Storing and Loading Instrument Data - File Key
1407.0806.32 4.21 E-2
Save Immediate x
Stores the current instrument setting in one of the three intermediate
memories.
These instrument settings are retained until a different instrument
setting is stored in the intermediate memory. When the instrument is
switched off, the contents of the intermediate memories are retained.
Remote-control command:
*SAV 1
File Management
Calls the File Management menu.
Directories can be created and files managed in this menu (see the
section "File Management - File", page 4.23).
Remote-control command: n.a.
Loading Instrument Settings - File
If Recall is selected under Select Operation, the File menu provides options for loading complete
instrument settings. Here it is possible to select whether the current or stored RF frequency and RF
level settings are to be used.
Recent data sets - File
Displays the files last used. The entire path is shown.
If the desired file is in the list, it can be selected in this window.
Remote-control command: n.a.
Directory - File
Enter the directory in which the file with the instrument setting to be
loaded is located.
The Selected file window lists all the files in this directory.
Remote-control command:
MMEM:CDIR "/var/user"
File List - File
Selects the file with the desired instrument configuration.
Remote-control command:
MMEM:CAT?