EasyManua.ls Logo

Rohde & Schwarz SMB 100a - Status Reporting System; Overview of the Status Register

Rohde & Schwarz SMB 100a
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S SMB Status Reporting System
1407.0806.32 5.23 E-2
Status Reporting System
The status reporting system stores information on errors which have occurred. This information is
stored in theerror queue. The error queue can be queried via IEC/IEEE bus or via the Ethernet.
The information is of a hierarchical structure. The register status byte (STB) defined in IEEE 488.2 and
its associated mask register service request enable (SRE) form the uppermost level. The STB receives
its information from the standard event status register (ESR) which is also defined in IEEE 488.2 with
the associated mask register standard event status enable (ESE).
The IST flag ("I
ndividual STatus") and the parallel poll enable register (PPE) allocated to it are also part
of the status reporting system. The IST flag, like the SRQ, combines the entire instrument status in a
single bit. The PPE fulfills an analog function for the IST flag as the SRE for the service request.
The output buffer contains the messages the instrument returns to the controller. It is not part of the
status reporting system but determines the value of the MAV bit in the STB.
Overview of the Status Register
The following figure shows the status registers used in the Signal Generator.
SRE
STB
PPE
IST flag
(response to parallel poll)
& = logical AND
= logical OR
of all bits
ESE ESR
Error Queue
Output Buffer
SRQ
RQS/MSS
ESB
MAV
Power on
User Request
Command Error
Execution Error
Device Dependent Error
Query Error
Request Control
Operation Complete
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
7
6
5
4
3
2
1
0
Overview of status registers

Table of Contents

Related product manuals