R&S SMB Alphabetical List of SCPI-Error Messages
1407.0806.32 9.9 E-2
Settings conflict ... (-221)
There is a setting conflict between the two parameters indicated after the semicolon.
Example:
The set FM deviation is too large for the selected RF frequency.
Remedy
One of the given values has to be corrected to obtain a valid
output signal.
SCPI:
Execution error - sets bit 4 in the ESR register
String data not allowed (-158)
The command contains a legal string data element which is not allowed at this point.
Example:
A text parameter is set in quotation marks -
SOURce:FREQuency:MODE "FIXed".
SCPI:
Command error - sets bit 5 in the ESR register
Suffix not allowed (-138)
A suffix is not allowed for this command or at this point of the command.
Example:
Command *RCL does not permit indicating a suffix.
SCPI:
Command error - sets bit 5 in the ESR register
Suffix too long (-134)
The suffix contains more than 12 characters.
SCPI:
Command error - sets bit 5 in the ESR register
Syntax error (-102)
The command is invalid.
Example:
The command contains block data the instrument does not
accept.
SCPI:
Command error - sets bit 5 in the ESR register
System error (-310)
This error message suggests an error within the instrument. Please inform the R&S
Service.
SCPI:
Device specific error- sets bit 3 in the ESR register
Too many digits (-124)
The decimal numeric data element contains too many digits.
SCPI:
Command error - sets bit 5 in the ESR register