R&S SMB Alphabetical List of SCPI-Error Messages
1407.0806.32 9.7 E-2
Invalid suffix (-131)
The suffix is not appropriate for this command.
Example:
nHz is not defined.
SCPI:
Command error - sets bit 5 in the ESR register
Lists not same length (-226)
The parts of a list have different lengths. This error message is also displayed if only
part of a list has been transmitted via IEC bus. All parts of the list have to be
transmitted always before it is executed.
Example:
The POWer list content is longer than the FREQuency list
content, or only the POWer content is transmitted.
SCPI:
Execution error - sets bit 4 in the ESR register
Missing parameter (-109)
The command does not contain the required parameters.
Example:
Command :SOUR:FREQ requires the indication of a parameter -
":FREQ; POW -35"
SCPI:
Command error - sets bit 5 in the ESR register
No error (0)
This message is output if the error queue does not contain entries.
Numeric data not allowed (-128)
The command contains a numeric data element the device does not accept in this
position.
Example:
The command :SOUR:BB:MCCW:SEQ requires the indication of a
text parameter - SOUR:BB:MCCW:SEQ AUTO.
SCPI:
Command error - sets bit 5 in the ESR register
Out of memory (-225)
The storage space available in the instrument is exhausted.
SCPI:
Execution error - sets bit 4 in the ESR register
Parameter not allowed (-108)
The command contains too many parameters.
Example:
Command SOURce:FM:INTernal:FREQuency permits only
one frequency indication - ":FREQ:CENT 30 kHz, 40 kHz".
SCPI:
Command error - sets bit 5 in the ESR register