General Instrument Settings R&S SMB
1407.0806.32 4.16 E-2
Layout
Selects the keyboard layout for the selected keyboard language. The
assignment of some keys depends on the selected layout and
language.
Remote-control command:
KBO:LAY US
Language
Selects the keyboard language. The assignment of some keys
depends on the selected language.
Remote-control command::
KBO:LANG US
Remote Channel Settings - Setup-Remote
The Remote GPIB/Ethernet... menu provides access to the GPIB and Ethernet settings. The Remote
GPIB... menu is opened using the SETUP or MENU key under Remote.
GPIB channel address
Sets the IEC/IEEE-bus address of the instrument.
Remote-control command:
SYST:COMM:GPIB:ADDR 28
Visa Resource String
Indicates the visa resource string. This string is used for remote
control of the instrument. A separate string is provided for remote
control via the LAN and the USB interface
Remote-control commands:
SYST:COMM:NETW:RES?
Response: TCPIP::192.1.2.3::INSTR
SYST:COMM:USB:RES?
Response: USB::72::000000::INSTR
Protection - Setup-Protection
The Protection menu provides access to the unlocking of protected service functions (authorized
personnel of R&S Service Departments only). To deactivate the protection, the correct password has to
be entered. After the instrument has been switched on, the protection levels 1 to 5 are automatically
activated.