R&S SMB RF Signal - RF
1407.0806.32 4.27 E-2
Frequency - Menu
The Frequency menu is opened in the RF function block or using the MENU key under RF.
Frequency and offset of the RF output signal is set in the upper part of the group menu. The step width
which is used when setting the frequency using the rotary knob (with Variation Active On) is set in the
User Variation section.
The lower parts concern the phase settings for the output signal and are described in the section "RF
Phase", page 4.29.
The frequency and offset are set in the top section of the menu.
Frequency - RF Signal
Sets the RF frequency of the RF output connector. The frequency
entered and displayed here corresponds to the frequency at the RF
output, i.e. any offset entry is not taken into consideration.
Remote-control command: n.a.
Note:
There is no GPIB command available for this setting. The GPIB
command SOUR:FREQ sets the frequency of the FREQ display, i.e.
the frequency containing offset.
Frequency Offset - RF
Signal
Sets the frequency offset relative to the RF frequency. The frequency
offset of a downstream instrument (e.g. a mixer) is entered.
The entry does not change the RF frequency at the RF output. It only
changes the RF frequency displayed in the display header. the RF
frequency in the header corresponds to the frequency.
Remote-control command:
SOUR:FREQ:OFFS 0 Hz
If the frequency is set using the rotary knob, the step width is defined in the User Variation section.