R&S SMB RF Level
1407.0806.32 4.35 E-2
Attenuator Fixed Range In
– RF Level
Displays the level range in which the level is set without interruption
for the ā€œAttenuator Mode fixed" setting.
Remote-control commands:
OUTP:AFIX:RANG:UPP?
OUTP:AFIX:RANG:LOW?
If the level is set using the rotary knob, the step width is defined in the User Variation section.
Variation Step - RF Level
Sets the user-defined step width for entering the RF level using the
rotary knob. Level variation with this step width must also be activated
with Variation Active.
Remote-control command:
SOUR:POW:STEP:INCR 1dBm
Variation Active - RF Level
Activates the user-defined step width used when varying the level
value with the rotary knob.
ON
The level value set with the rotary knob is varied
using the user-defined step width which is entered
under Variation Step.
Remote-control command:
SOUR:POW:STEP:MODE USER
OFF
The level value set with the rotary knob is varied in
steps of one unit at the cursor position (standard
operating mode).
Remote-control command:
SOUR:POW:STEP:MODE DEC
Overvoltage Protection
The overvoltage protection prevent overloading by an external signal applied to the RF output of the
R&S SMB Signal Generator.
The overload protection is tripped when the power of the external signal becomes too high. A relay
opens and interrupts the connection between the RF output and attenuator. This condition is indicated
in the display header by the 'OVERLOAD' status message.
! Reset the overload protection by pressing the
RF ON/OFF key. The RF input is activated when the
overload protection is reset.
Remote-control commands:
OUTP:PROT:TRIP?
OUTP:PROT:CLE
OUTP:STAT ON