General Instrument Settings R&S SMB
1407.0806.32 4.8 E-2
State Test Point
Activates the measurement of the voltage at the selected test point.
Remote-control command:
DIAG1:MEAS:POIN? 'DIAG_NO_DP'
Response: 2
Select Test Point
Calls the submenu for selecting the test point. The currently selected
key is shown next to the key.
Remote-control commands:
DIAG1:POIN:CAT? (Command lists all test points)
DIAG1:POIN? 'DIAG_SMB_BAB_P7V'
(With remote control, voltage measurement starts as
soon as the test point is selected).
Voltage Test Point
Displays the measured voltage measurement at the selected test
point.
Remote-control command:
DIAG1:MEAS:POIN? 'DIAG_SMB_BAB_P7V'
Response: 2
Check Front Panel - Setup-Test
The Check Front Panel menu.. is used to check whether the front panel keys are working correctly.
The menu displays all the front panel keys arranged in the same way as on the front panel. The
respective function is executed by pressing the associated key.