General Instrument Settings R&S SMB
1407.0806.32 4.4 E-2
Internal Adjustments - Setup-System
The R&S SMB Signal Generator is extremely accurate thanks to the integrated procedures for
adjustments.
All internal adjustments for which no external measuring equipment is needed can be started in the
Internal Adjustments... menu. The adjustments with external measuring equipment are described in
the Service Manual (on CD ROM, supplied with the instrument).
Adjustment is recommended if the temperature range in which the instrument is operated changes, or
prior to all applications which require maximum level and frequency accuracy.
During adjustment a bar indicates the status of progress. If an error occurs, adjustment is terminated
and an error message is output in the info line.
ATTENTION
Risk of invalid adjustment
In order to achieve correct adjustment of the instrument, make sure that the
instrument is warm before performing adjustments. The warm-up time is 30 minutes.
The Internal Adjustments menu is opened up using the SETUP or MENU key under System.
Adjust All
Starts all internal adjustments for which no external measuring
equipment is needed. The adjustments with external measuring
equipment are described in the Service Manual (supplied).
Remote-control command:
CAL:ALL?
Adjust Synthesis
Performs all adjustments which affect the frequency.
Remote-control command:
CAL:FREQ:MEAS?
Level
Performs all adjustments which affect the level. The aquired correction
values improve the settling time and the signal quality.
Remote-control command:
CAL:LEV:MEAS?
Use ext. Level Correction
Data
Switch on or off use of external level correction data (see service
manual, chapter 2, "Adjustment"). This switch can be used in case of
corrupt external level correction data.
Remote-control command:
CAL:LEV:STAT ON