Loading...
Sweep Mode R&S SMB
1407.0806.32 4.48 E-2
Reset Sweep – RF
Frequency Sweep
Resets the sweep. The start frequency is set and the next sweep
starts from there.
Remote-control command:
SWE:RES:ALL
Start Freq - RF Frequency
Sweep
Sets the start frequency.
Remote-control command:
SOUR:FREQ:STAR 100MHz
Stop Freq - RF Frequency
Sweep
Sets the stop frequency.
Remote-control command:
SOUR:FREQ:STOP 500MHz
Center Freq - RF
Frequency Sweep
Sets the center frequency.
Remote-control command:
SOUR:FREQ:CENT 300MHz
Span - RF Frequency
Sweep
Sets the span.
Remote-control command:
SOUR:FREQ:SPAN 400MHz
Current Freq - RF
Frequency Sweep
Displays the current frequency.
If Step is set, the frequency for the next frequency step of the sweep
is entered here.
Remote-control command
SOUR:FREQ:MAN 300MHz
Spacing - RF Frequency
Sweep
Selects linear or logarithmic sweep spacing.
Remote-control command:
SOUR:SWE:FREQ:SPAC LIN | LOG
Step Lin/Log - RF
Frequency Sweep
Sets the step width for the individual sweep steps. This entry is
effective for all sweep modes.
Step Lin or Step Log is displayed depending on whether Spacing
Lin or Log is selected.
Step Lin
With the linear sweep, the step width is a fixed
frequency value which is added to the current
frequency. The linear step width is entered in Hz.
Remote-control command:
SOUR:SWE:FREQ:STEP:LIN 1 MHz