R&S SMB Sweep Mode
1407.0806.32 4.51 E-2
Mode - RF Level Sweep
Selects the Level Sweep instrument operating mode and the Sweep
mode.
Auto
Sets an automatic repeated sweep cycle. If a
different sweep mode was activated prior to the
Auto mode, the cycle continues from the current
sweep setting.
The Reset button resets the sweep to the start level.
Remote-control commands:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR AUTO
SOUR:POW:MODE SWE
Single
Sets a single sweep cycle. The sweep is triggered
by the Execute Single Sweep button.
If a different sweep mode was activated prior to the
Single mode, the current sweep is stopped. The
Single sweep always starts at the start level.
The Reset button stops the sweep and resets it to
the start level.
Remote-control commands:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:POW:EXEC
Step
Sets a step-by-step sweep cycle.
If this mode is activated, the cursor moves to the
value displayed for Current Level. Each sweep step
is triggered by a variation of the value in the Current
Level entry window. The step width is set below at
entry field Step.
If this mode is activated, the cursor moves to the
value displayed for Current Level. If a different
sweep mode was activated prior to the Step mode,
the current sweep is stopped. The Step sweep
starts at the current level value.
The Reset button resets the sweep to the start level.
Remote-control commands:
SOUR:SWE:POW:MODE MAN
SOUR:POW:MODE SWE
SOUR:SWE:POW:STEP 0.5
SOUR:POW:MAN -15
(the value entered with command
SOUR:SWE:POW:STEP sets the step width. The
value entered with command SOUR:POW:MAN has
no effect, the command only triggers the next sweep
step. However, the value has to be in the sweep
range (start to stop value) In remote control only a
step-by-step sweep from start to stop level is
possible)