R&S SMB List Mode
1407.0806.32 4.59 E-2
Save as
Open the file menu to save the list under a new
name.
Each list is saved to the R&S SMA Signal Generator
CompactFlashâ„¢ Card as a separate file with the file
prefix *.lsw. The file name and the directory to which
the file is saved are user-selectable.
Remote-control command: n.a.
Save
The list is saved under its current name.
Remote-control command: n.a.
(the list is saved automatically after the values have
been entered)
List Range In - List Mode
Defines an index range in the current list by setting the start and stop
index. Only the values in the selected index range are processed in
LIST mode, all other list entries are ignored.
Remote-control command:
SOUR:LIST:IND:STAR 15
SOUR:LIST:IND:STOP 155
Import/Export - List Mode
Expands the menu with the area for import and export of list mode
files.
Externally edited Excel tables with frequency/level pairs can be
imported as text or CSV-files and used for list mode. On the other
hand, internally created list mode list can be exported as text or CSV-
files..
Mode - List Mode
Selects if list mode lists should be imported or exported. . The
settings offered below depend on the selected mode.
Remote-control command:
SOUR:LIST:DEXC:MODE IMP
Extension - List Mode
Selects the file extension of the ASCII file to be imported or exported.
Selection TXT (text file) or CSV (Excel file) is available. .
Remote-control command:
SOUR:LIST:DEXC:AFIL:EXT TXT
Decimal Point - List Mode
Selects the decimal separator used in the ASCII data between '.'
(decimal point) and ',' (comma) with floating-point numerals. .
Remote-control command:
SOUR:LIST:DEXC:AFIL:SEP:DEC DOT
Column Separator- List
Mode
Selects the separator between the frequency and level column of the
ASCII table.
Remote-control command:
SOUR:LIST:DEXC:AFIL:SEP:COL TAB