EasyManua.ls Logo

Rohde & Schwarz SMB 100a - Page 155

Rohde & Schwarz SMB 100a
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S SMB LF Generator and LF Output - Mod Gen Block
1407.0806.32 4.77 E-2
State - LF Sweep
Activates LF Sweep mode.
Note:
Activating the LF sweep automatically deactivates the RF
frequency sweep and level sweep.
Remote-control commands:
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:FREQ:MODE CW
Mode - LF Sweep
Selects the Sweep instrument operating mode and Sweep mode.
Auto
Sets an automatic repeated sweep cycle. If a
different sweep mode was activated prior to the
Auto mode, the cycle continues from the current
sweep setting.
The Reset button resets the sweep to the start
frequency.
Remote-control commands:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR AUTO
SOUR:LFO:FREQ:MODE SWE
Single
Sets a single sweep cycle. The sweep is triggered
by the Execute Single Sweep button.
If a different sweep mode was activated prior to the
Single mode, the current sweep is stopped. The
Single sweep always starts at the start frequency.
The Reset button stops the sweep and resets it to
the start frequency.
Remote-control commands:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR SING
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:SWE:FREQ:EXEC
Step
Sets a step-by-step sweep cycle. Each sweep step
is triggered by a variation of the value in the Current
Freq entry window.
If this mode is activated, the cursor moves to the
value displayed for Current Freq. If a different
sweep mode was activated prior to the Step mode,
the current sweep is stopped. The Step sweep
starts at the current LF frequency.
The Reset button resets the sweep to the start
frequency.
Remote-control commands:
SOUR:LFO:SWE:FREQ:MODE MAN
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:SWE:FREQ:SPAC LIN
SOUR:LFO:SWE:FREQ:STEP:LIN 1E3
SOUR:LFO:FREQ:MAN 12 kHz