R&S SMB Switchover to Remote Control
1407.0806.32 5.5 E-2
! Select TCP/IP Resource and open the next page of the Create new menu with the right mouse key.
! Enter the IP address of the Signal Generator and select VXI-11 Instrument
(the computer name may be entered instead).
! Press the Test button
A message indicates whether the link to the Signal Generator can be set up or not. If a connection
cannot be set up, check whether the controller and the instrument are connected to the network
(network cable) and switched on. Correct spelling of the IP address or the computer name can also
be checked. For further error location, inform the network administrator. In large networks,
specification of additional addresses may be required for link setup, e.g. gateway and subnet mask,
which are known to the network administrator.