SCPI Command Structure and Syntax R&S SMB
1407.0806.32 5.12 E-2
Example:
[SOURce]:POWer[:LEVel][:IMMediate]:OFFSet 1
This command immediately sets the offset of the signal to 1 dB. The following
command has the same effect:
POWer:OFFSet 1
Long and short form: The key words feature a long form and a short form. Either the short form or
the long form can be entered, other abbreviations are not permitted.
Example: STATus:QUEStionable:ENABle 1= STAT:QUES:ENAB 1
Note:
Upper-case and lower-case notation only serves to distinguish the two
forms in the manual, the instrument itself does not distinguish upper-case
and lower-case letters.
Parameters:
Parameters must be separated from the header by a "white space". If several
parameters are specified in a command, they are separated by a comma ",".
A few queries permit the parameters MINimum, MAXimum and DEFault to be
entered. For a description of the types of parameter, refer to Section
"Parameters".
Example: SOURce:POWer:ATTenuation? MAXimum Answer: 60
his query requests the maximal value for the attenuation.
Numeric suffix:
If a device features several functions or features of the same kind, e.g.
several inputs, the desired function can be selected by a suffix added to the
command. Entries without suffix are interpreted like entries with the suffix 1.
Optional keywords must be specified if they select a function with the suffix.
Example:
SOURce:LFOutput2:FREQuency 1.5kHz
This command sets the frequency of secondLF generator frame.