User Correction R&S SMC100A
1411.4060.32 4.60 E-1
The combined menu ALC / UCOR menu provides access to function User Correction in the User
Correction Settings section. The menu is used to activate/deactivate user correction, and to create,
select and activate the correction data lists.
The upper sections provide access to the automatic level control settings, see section "Automatic Level
Control - ALC", page 4.57 .
State - User Correction
Activates/deactivates user correction.
The UCOR status message appears in the frequency and level
display.
Remote-control command:
SOUR:CORR:STAT ON
Value - User
Correction
Indicates the current value for level correction.
Remote-control command:
SOUR:CORR:VAL?
User Correction Data -
User Correction
Calls the File Select menu for selecting and creating a list or the File
Manager.
Remote-control command:
MMEM:CDIR '/var/ucor'
SOUR:CORR:CSET:CAT?
Response:
'Ucor1', 'Ucor2'
SOUR:CORR:CSET:SEL 'Ucor1'
SOUR:CORR:CSET:DEL 'Ucor2'