R&S SMC100A User Correction
1411.4060.32 4.61 E-4
Edit/New User
Correction Data - User
Correction
Calls the editor for editing the selected user correction list. A list
consists of any number of frequency/level value pairs. The currently
selected list is displayed.
Each list is saved to the R&S SMC Signal Generator as a separate file
with the file prefix *.uco. The file name and the directory to which the
file is saved are user-selectable.
Note:
Save a list only after filling both columns (frequency and level),
otherwise the entries are lost.
Frequency /Hz Enters the frequency to which the level correction
value applies.
Note:
The Fill.. function allows to automatically enter
any number of frequencies with freeley selectable
range and increment.
Using the Fill With Sensor function of the Edit
sub menu requires only the entry of the fre-
quency values. The level values are automati-
cally acquired by the connected power sensor.
Remote-control command:
SOUR:CORR:CSET:SEL 'Ucor1'
SOUR:CORR:CSET:DATA:FREQ 100MHz,...
Power /dB
Enters the level correction value to which the
specified frequency applies. The values can be
entered manually or automatically with the Fill With
Sensor function.
Remote-control command:
SOUR:CORR:CSET:SEL 'Ucor1'
SOUR:CORR:CSET:DATA:POW 1dB,0.8dB,...
Goto
Selects the row for editing.
If Goto Row... is selected, a window opens for
entering the requested row.
Remote-control command: n.a.
(it is not possible to change individual positions of
the list)