R&S SMC100A Sweep Mode
1411.4060.32 4.77 E-4
Extern
Start/Stop
Sets an automatically repeated sweep cycle that is
started, stopped and restartet by subsequent
external trigger events.
The first external trigger signal starts the sweep
(Start).
The next external trigger signal stops the sweep at
the current level (Stop).
The third external trigger signal starts the sweep at
the start level (Start).
If a different sweep mode was activated prior to the
Extern Start/Stop mode, the current sweep is
stopped and the Extern Start/Stop sweep starts at
the start level when triggered.
The Reset button resets the sweep to the start level.
The external trigger signal is input at the rear of the
instrument (BNC connector INST TRIG).
Remote-control command:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EAUT
SOUR:POW:MODE SWE
(External trigger)
Reset Sweep - RF
Level Sweep
Resets the sweep. The start level is set and the next sweep starts
from there.
Remote-control command:
SWE:RES:ALL
Execute Single Sweep
- RF Level Sweep
Triggers the sweep manually. A manual sweep can only be triggered if
Mode Single is selected.
Remote-control commands:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:POW:EXEC