Command Description R&S SMC100A
1411.4060.32 6.34 E-1
[SOURce:]CORRection:CSET:CATalog?
The command requests a list of user correction tables. The individual lists are separated by
commas.
The lists are stored with the fixed file extensions *.uco in a directory of the user's choice. The
directory applicable to the commands is defined with the command MMEMory:CDIR.
This command is a query command and has no *RST value.
Example:
MMEM:CDIR ‘var/lists/ucor’
'selects the directory for the user correction files.
CORR:CSET:CAT?
'queries which correction tables are available.
Response: UCOR1,UCOR2,UCOR3
'the correction tables UCOR1, UCOR2 and UCOR3 are available.
*RST value Resolution Options SCPI
- - - Device-specific
[SOURce:]CORRection:CSET:DATA:FREQuency 300 kHz...RF
max
{,300 kHz...RF
max
}
The command transfers the frequency data to the table selected with :CORR:CSET:SEL.
*RST does not affect data lists.
Example:
CORR:CSET ‘var/lists/ucor/ucor1'
'selects the table ucor1.
CORR:CSET:DATA:FREQ 100MHz,102MHz,103MHz,...
'enters the frequency value in the table UCOR1.
*RST value Resolution Options SCPI
- - - Device-specific
[SOURce:]CORRection:CSET:DATA:FREQuency:POINts?
The command queries the number of frequency values in the selected table.
The command is a query command and therefore has no *RST value.
Example:
CORR:CSET ‘var/lists/ucor/ucor1'
'selects the table ucor1.
CORR:CSET:DATA:FREQ:POIN?
'queries the number of frequency values in the table ucor1.
Response: 440
'the table ucor1 contains 440 frequency values.
*RST value Resolution Options SCPI
- - - Device-specific