R&S SMC100A Command Description
1411.4060.32 6.35 E-1
[SOURce:]CORRection:CSET:DATA:POWer -40dB ... 6dB {,-40dB ... 6dB}
The command transfers the level data to the table selected with :CORRection:CSET:SELect.
*RST does not affect data lists. .
Example:
CORR:CSET ‘var/lists/ucor/ucor1'
'selects the table ucor1.
CORR:CSET:DATA:POW 1dB, 0.8dB, 0.75dB,...
'enters the level values in the table ucor1.
*RST value Resolution Options SCPI
- - - Device-specific
[SOURce:]CORRection:CSET:DATA:POWer:POINts?
The command queries the number of level values in the selected table.
The command is a query command and therefore has no *RST value.
Example:
CORR:CSET ‘var/lists/ucor/ucor1'
'selects the table ucor1.
CORR:CSET:DATA:POW:POIN?
'queries the number of level values in the table ucor1.
Response: 440
'the table ucor1 contains 440 level values.
*RST value Resolution Options SCPI
- - - Device-specific
[SOURce:]CORRection:CSET:DATA[:SENSor][:POWer]:SONCe
The command activates the filling of the user correction list with level values acquired by the power
sensor.
The command activates the filling of the user correction list with level values acquired by the
selected power sensor. All level correction values for the given frequency values are entered in the
selected list.
The command is an event and therefore has no *RST value and no query.
Example:
"CORR:CSET:DATA:SENS:POW:SONC"
' fills the user correction list with level values acquired by the power sensor.
*RST value Resolution SCPI
- - Device-specific