Command Description R&S SMC100A
1411.4060.32 6.74 E-1
4. Select the trigger mode.
TRIGger:FSWeep:SOURce SINGle
5. Select the sweep mode and activate the sweep.
[
SOURce:]SWEep[:FREQuency]:MODE AUTO
[SOURce:]FREQuency:MODE SWEep
6. Trigger the sweep.
[SOURce:]SWEep[:FREQuency]:EXECute
Note:
It is recommended to switch off the GUI update for optimum sweep performance especially with
short dwell times (SYSTem:DISPlay:UPDate OFF).
Command Parameters Default
unit
Remark
[SOURce:]SWEep[:FREQuency]:DWELl 2 ms ... 10 s s
[SOURce:]SWEep[:FREQuency]:EXECute -
[SOURce:]SWEep[:FREQuency]:MODE AUTO | MANual | STEP
[SOURce:]SWEep[FREQuency]:POINts <numeric_value>
[SOURce:]SWEep[:FREQuency]:SHAPe SAWTooth |TRIangle
[SOURce:]SWEep[:FREQuency]:SPACing LINear | LOGarithmic
[SOURce:]SWEep[:FREQuency]:STEP[:LINear] 0 ... (STOP - STARt) Hz
[SOURce:]SWEep[:FREQuency]:STEP:LOGarithmic 0.01 ... 9999 PCT
[SOURce:]SWEep:POWer:DWELl 2 ms...10 s s
[SOURce:]SWEep:POWer:EXECute -
[SOURce:]SWEep:POWer:MODE AUTO | MANual | STEP
[SOURce:]SWEep:POWer:POINts <numeric_value>
[SOURce:]SWEep:POWer:SHAPe SAWTooth |TRIangle
[SOURce:]SWEep:POWer:SPACing:MODE
Query only
[SOURce:]SWEep:POWer:STEP:[:LOGarithmic] 0.1 ...20 dB
[SOURce:]SWEep:RESet[:ALL]
No query

Table of Contents