R&S SMC100A Command Description
1411.4060.32 6.75 E-1
[SOURce:]SWEep[:FREQuency]:DWELl 3 ms...10 s
The command sets the time taken for each frequency step of the sweep.
Note:
It is recommended to switch off the GUI update for optimum sweep performance especially with
s
hort dwell times (SYSTem:DISPlay:UPDate OFF).
The keyword [:FREQuency] can be omitted (see example). The command is then SCPI-compliant.
Example:
SWE:DWEL 12 ms
'sets a dwell time of 12 ms for a frequency sweep.
*RST value Resolution Options SCPI
10 ms 0.1 ms Device-specific
[SOURce:]SWEep[:FREQuency]:EXECute
The command triggers a sweep. The command is only valid for sweep mode Single
(SOURce:SWEep:FREQuency:MODE SINGle). The command corresponds to the manual-control
command EXECUTE SINGLE SWEEP.
The command triggers an event and therefore has no query form and no *RST value.
Example:
SWE:FREQ:MODE SING
'sets the single cycle mode of the frequency sweep.
SWE:FREQ:EXEC
'triggers a single frequency sweep.
*RST value Resolution Options SCPI
- - Device-specific