R&S SMC100A Example of Setup
1411.4077.62 2.11 E-1
Step 3: Activate RF signal
Activate the RF signal output by pressing the
RF ON/OFF
key.
T
o indicate the active state, the RF block is
displayed in blue. An RF signal with the default
frequency and level settings is output, i.e. 1 GHz
and -30 dBm.
The single sweep still must be triggered in the
sweep menu.
Step 4: Trigger single RF frequency sweep
Select the Execute Single Sweep button in the
RF frequency sweep menu by turning the rotary
knob. Press the rotary knob.
A linear single sweep from 100 MHz to 500 MHz in 1 MHz steps is output with a dwell time of 10 ms per
step.