General Instrument Settings R&S SMC100A
1411.4060.32 4.18 E-1
Mode
Selects if the IP address is assigned automatically or manually or if a
Peer to Peer connection is used.
The selection is only accepted after the Accept button is pressed.
A
uto (DHCP)
The IP address is assigned automatically.
The network used must support automatic
assignment of IP address (DHCP) in order to use
this function.
Remote-control command:
SYST:COMM:NETW:IPAD:MODE AUTO
Static
The IP address is assigned manually.
Remote-control command:
SYST:COMM:NETW:IPAD:MODE STAT
Peer-to-Peer
A Peer to Peer connection is used.
Remote-control command:
SYST:COMM:NETW:IPAD:MODE P2P
IP Address
Only for Static and Peer to Peer mode.
Enters the IP address manually.
The new entry is only accepted and the IP address changed after the
Accept button is pressed.
Remote-control command:
SYST:COMM:NETW:IPAD 7.8.9.10
Subnet Mask
Only for Static and Peer to Peer mode.
Enters the Subnet mask.
This number is used together with the IP address to identify the
network segment the instrument is in.
The new entry is only accepted and the subnet mask changed after
the Accept button is pressed.
Remote-control command:
SYST:COMM:NETW:IPAD:SUBN:MASK 255.255.255.0
Def Gateway
Only for Static mode.
Enters the IP address of the default gateway.
The new entry is only accepted and the default gateway is changed
after the Accept button is pressed.
This address identifies the router on the same network as the
instrument that is used to forward ttraffic to destinations beyond the
local network.
Remote-control command:
SYST:COMM:IPAD:NETW:GAT 456