EasyManua.ls Logo

Santec TSL-710 - Page 106

Default Icon
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-52
TSL-710 Operation Manual
SO Syntax SO
Description SetsOPENstatusoftheinternalshutter.
Parameter None
Response None
PR Syntax PR
Description Revisesthecompensatepowerdataofsweepsinautopowermode.
Recommendexecutingthiscommandwhen"OverLD"or"LowPwr"
isdisplayed.
Parameter None
Response None
SS Syntax SS<value>
Description Setsthesweepstartwavelength.Whenparameterisnotspecied,reads
outthecurrentlysetvalue.
Parameter Range: MinimumsettingwavelengthtoMaximum(nm)
Step: 0.0001(nm)
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setsthesweepstartwavelengthto1510nm.
SS1510
Readsoutthecurrentlysetvalue.
SS
1560.000
Indicatesthatthecurrentlysetstartwavelengthis1560nm.
FS Syntax FS<value>
Description Setsthesweepstartfrequency.Whenparameterisnotspecied,reads
outthecurrentlysetvalue.
Parameter Range: MinimumsettingfrequencytoMaximum(THz)
Step: 0.00001(THz)
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setsthesweepstartfrequencyto199.8616THz.
FS199.8616
Readsoutthecurrentlysetvalue.
FS
199.8616
Indicatesthatthecurrentlysetstartfrequencyis199.8616THz.

Table of Contents