7-54
TSL-710 Operation Manual
SB Syntax SB<value>
Description Setstheamountoftimespentduringeachstepinstepsweepoperation.
Thewaittimebetweeneachsweepisalsosettothesamevalue.This
waittimedoesnotincludethetimerequiredforwavelengthconversion.
Whenparameterisnotspecied,readsoutthecurrentlysetvalue.
Parameter Range:0~999.9(s)
Step: 0.1(s)
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setsthewaittimeofeachstepto1second.
SB1
Readsoutthecurrentlysetvalue.
SB
0000.50
Indicatesthatthecurrentlysetwaittimeofeachstepis0.5
second.
SZ Syntax SZ<value>
Description Setsthenumberofwavelengthsweeps.Whenparameterisnotspecied,
readsoutthecurrentlysetvalue.
Parameter Range: 0 ~999
Step: 1
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setsthenumberofsweepsto100.
SZ100
Readsoutthecurrentlysetvalue.
SZ
000100
Indicatesthatthenumberofsweepsissetto100.
SN Syntax SN<value>
Description Setsthespeedforcontinuoussweep.Whenparameterisnotspecied,
readsoutthecurrentlysetvalue.
Parameter Range: 0.5~100(nm/s)
Step: 0.1(nm/s)
Response Currentlysetvalue.Avalueisreturnedbothonsettingandonreading.
Example Setsthespeedforcontinuoussweepto100nm/s.
SN100
Readsoutthecurrentlysetvalue.
SN
100.000
Indicatesthatthesweepspeedissetto100nm/s.