7-26
TSL-710 Operation Manual
5. TSL-710 specic command detailed description
Specic commands in Command set 1 are described in detail as follows.
1) System related commands
:SPECial:REBoot
Syntax :SPECial:REBoot
Description Resets the device.
Parameter None
Response None
Example :SPEC:REB
:SYSTem:ERRor?
Syntax :SYSTem:ERRor?
Description Reads out the error issued.
Parameter None
Response Error Number (Refer to “6. Command Error” below.)
Example :SYST:ERR?
Error numbers and error messages are returned.
:SYSTem:VERSion?
Syntax :SYSTem:VERSion?
Description Reads out the version of the rmware.
Parameter None
Response The format is (4 digits) + (period) + (4 digits)
****.****
Example :SYST:VERS?
0001.0000
:SYSTem:COMMunicate:GPIB:ADDRess
Syntax :SYSTem:COMMunicate:GPIB:ADDRess
<wsp><value>
Description Sets the GPIB address.
Parameter Integer from 1 to 30
Response None
Example :SYST:COMM:GPIB:ADDR 10
Sets the GPIB address to “10”.
:SYSTem:COMMunicate:GPIB:ADDRess?
Syntax :SYSTem:COMMunicate:GPIB:ADDRess?
Description Reads out the GPIB address.