Tunable Semiconductor Laser TSL-710
7-35
Parameter None
Response Range: 30dB up to peak power
Typical value
-20 ~ +10dBm (0.01 ~ 10mW)
The available range of output power
depends on the product characteristics.
Step: 0.01dBm (0.01mW)
The unit is previously set with
the [:SOURce]:POWer:UNIT command.
Example : POW: ACT?
9.600
The output power is 9.6 dBm or 9.6 mW
depending on the value set for the unit.
[:SOURce]:POWer:CORRection:PROFile
Syntax [:SOURce]:POWer:CORRection:PROFile
Description Revises the compensate power data of sweeps in
auto power mode.
Recommend executing this command when
“Over LD”or “Low Pwr” is displayed.
Parameter None
Response None
Example :POW:CORR:PROF
[:SOURce]:WAVelength:UNIT
Syntax [:SOURce]: WAVelength: UNIT <wsp> <value>
Description Sets the unit of the wavelength indication
(Wavelength, Frequency).
Separate commands exist for setting
frequency and wavelength.
Parameter  0nm
1THz
Response None
Example :WAV:UNIT 1
Sets the wavelength unit t THz.
[:SOURce]:WAVelength:UNIT?
Syntax [:SOURce]:WAVelength:UNIT?
Description  Reads out the unit setting of the
wavelength indication.
Parameter None
Response 0nm
1THz