EasyManua.ls Logo

Santec TSL-710 - Page 90

Default Icon
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-36
TSL-710 Operation Manual
Example :WAV:UNIT?
1
Indicates that the wavelength unit is THz.
[:SOURce]:WAVelength
Syntax [:SOURce]:WAVelength<wsp><value>
Description  Sets the output wavelength.
Parameter RangeRange: Minimum setting wavelength
to Maximum (nm)
Step0.0001 (nm)
Response None
Example :WAV 1550
Sets the output wavelength to 1550 nm.
[:SOURce]:WAVelength?
Syntax [:SOURce]:WAVelength[:MINimum|:MAXimum]?
Description  Reads out the setting value of the wavelength.
Parameter None: Reads out the currently set wavelength.
MINimum:
Reads out the minimum acceptable wavelength.
MAXimum:
Reads out the maximum acceptable wavelength.
Response RangeMinimum acceptable wavelength
to Maximum acceptable wavelength (nm)
Step0.0001 (nm)
Example :WAV?
1550.000
Indicates that the currently set wavelength is 1550 nm.
[:SOURce]:WAVelength:FINe
Syntax [:SOURce]: WAVelength:FINe <wsp> <value>
Description  Sets the ne-tuning.
Parameter Range-100.00 +100.00  (Unit: None)
Step0.01 (Unit: None)
Response None
Example :WAV:FIN 50
Sets the ne-tuning to 50.
[:SOURce]:WAVelength:FINe?
Syntax [:SOURce]:WAVelength:FINe?
Description  Reads out the setting value of the ne-tuning.
Parameter None

Table of Contents