Tunable Semiconductor Laser TSL-710
7-39
[:SOURce]:WAVelength:SWEep:DELay?
Syntax [:SOURce]:WAVelength:SWEep:DELay?
Description Reads out the wait time between each sweep in
continuous sweep operation.
Parameter None
Response Range:0 〜 999.9 (s)
Step:0.1 (s)
Example : WAV:SWE:DEL?
→ 1.0
Indicates that the wait time is set to 1 second.
[:SOURce]:WAVelength:SWEep:DWELl
Syntax
[:SOURce]:WAVelength:SWEep:DWELl <wsp> <value>
Description Sets the amount of time spent during each step
in step sweep operation. The wait time
between each sweep is also set to the
same value. This wait time does not include
the time required for wavelength conversion.
Parameter Range:0 〜 999.9 (s)
Step:0.1 (s)
Response None
Example :WAV:SWE:DWEL 1
Sets the wait time of each step to 1 second.
[:SOURce]:WAVelength:SWEep:DWELl?
Syntax [:SOURce]:WAVelength:SWEep:DWELl?
Description Reads out the amount of time spent during each
step in step sweep operation.
Parameter None
Response Range:0 〜 999.9 (s)
Step:0.1 (s)
Example :WAV:SWE:DWEL?
→ 3.0
Indicates that the wait time is set to 3 seconds.
[:SOURce]:WAVelength:SWEep:MODe
Syntax
[:SOURce]:WAVelength:SWEep:MODe <wsp> <value>
Description Sets the sweep operation mode.
Parameter 0: Step operation, One-way
1: Continuous operation, One-way
2: Step operation, Two-way
3: Continuous operation, Two-way