7-42
TSL-710 Operation Manual
[:SOURce]:WAVelength:SWEep:STEP[:WIDTh]
Syntax [:SOURce]: WAVelength:SWEep:STEP [:WIDTh ]
<wsp> <value>
Description  Sets the step interval (wavelength) of step sweeps.
Parameter Range0.0001 160 (nm)
Step0.0001 (nm)
Response None
Example :WAV:SWE:STEP 0.1
Sets the step interval to 0.1 nm.
[:SOURce]:WAVelength:SWEep:STEP[:WIDTh]?
Syntax [:SOURce]:WAVelength:SWEep:STEP[:WIDTh]?
Description  Reads out the step interval (wavelength) of
step sweeps.
Parameter None
Response Range0.0001 160 (nm)
Step0.0001 (nm)
Example :WAV:SWE:STEP?
0.100
Indicates that the step interval is 0.1 nm.
[:SOURce][:WAVelength]:FREQuency:SWEep:STARt
Syntax [:SOURce][:WAVelength]:FREQuency:SWEep:STARt
<wsp><value>
Description  Sets the sweep start frequency.
Parameter Range: Frequency tuning range (THz)
Step0.00001 (THz)
Response None
Example :FREQ:SWE:STAR 185.0570
Sets the sweep start frequency to 185.0570 THz.
[:SOURce][:WAVelength]:FREQuency:SWEep:STARt?
Syntax
[:SOURce][:WAVelength]:FREQuency:SWEep:STARt?
Description  Reads out the sweep start frequency.
Parameter None
Response Range: Frequency tuning range (THz)
Step0.00001 (THz)
Example :FREQ:SWE:STAR?
185.0570
Indicates that the sweep start frequency
is 185.0570 THz.