7-44
TSL-710 Operation Manual
[:SOURce]:WAVelength:SWEep[:STATe]
Syntax
[:SOURce]:WAVelength:SWEep[:STATe] <wsp> <value>
Description  Sets the sweep mode state.
Parameter 0: Stop. Stops the sweep. Available during sweep
operation, and pause.
1: Start. Executes a sweep one time. Available in
sweep stop state. Repeating operations
are executed with other commands.
2: Pause. Pauses the sweep. Available only during
sweep operation.
3: Resume. Resumes a sweep from the pause state.
Available only during pause.
Response None
Example :WAV:SWE 1
Starts a sweep.
[:SOURce]:WAVelength:SWEep[:STATe]?
Syntax [:SOURce]:WAVelength:SWEep[:STATe]?
Description  Reads out the current sweep mode state.
Parameter None
Response 0: Stop
1: Operation
2: Pause
3: Trigger standby
4: Returning to START wavelength
* When state is “4”, sweep “STOP” and “PAUSE”
command are not accepted.
Example :WAV:SWE?
1
Indicates that a sweep is in operation.
[:SOURce]:WAVelength:SWEep[:STATe]:REPeat
Syntax [:SOURce]:WAVelength:SWEep[:STATe]:REPeat
Description  Executes multiple sweeps.
Parameter None
Response None
Example :WAV:SWE:REP
Executes multiple sweeps.