Trace | Multi Trace Operating Instructions 101
Operation
If “Controller active” is selected, the built-in
device controller is switched on and the feeding
pump is activated. While the controller is active,
the pump cannot be switched off or operated
manually.
To disable the controller, the check mark must be
clicked. In this state, the controller is inactive, but
the feeding pump is still running at the rate
specified by the last controller on.
With “Pump off” the feeding pump is switched
off manually. You can then turn on again only by
the re-activation of the controller.
In order to avoid a too high influence of the
integral gain on the PID Controller at the
beginning, the integral gain can be reset to the
preset value by “Reset controller” at any time.
This command must then be confirmed with
“Ok”. This function has no effect on the Mini-
mum-Maximum Controller.
“Pump manually” is a function that can be used
for filling the feeding tube. In this operating
window, the pump rate can be adjusted and the
pump is started or stopped. By “Exit” or close the
window, the pump is automatically stopped.