102 BioPAT
®
Trace | Multi Trace Operating Instructions
Operation
The respective type of control can be selected
between PID Controller and Minimum-Maximum
Controller in the sub menu “Parameters” under
“Type of controller”.
8.3.9.1 PID Controller
After selecting the type of controller to PID
Controller in the submenu “Parameters” the
“Set-point Glucose” is registered for the control.
Moreover, here the control parameters
“proportional gain”, “derivative time” and
“integral time” can be entered. In the option
“control by” it can be selected if the control acts
to the integrated pump or an external pump via
the serial interface or analog output 3.