104 BioPAT
®
Trace | Multi Trace Operating Instructions
Operation
8.3.9.2 Minimum-Maximum Controller
After selecting the type of controller to Mini-
mum-Maximum Controller in the submenu
“Parameters” the “Minimum” and “Maximum”
limits are registered for the control.
The Minimum-Maximum Controller switches the
feeding pump if it is above or below the limits
respectively to the minimum flow rate, or maxi-
mum flow rate.
The minimum delivery rate and the maximum
flow rate of the pump can be specified with the
setting “controller – minimal output” and “con-
troller – maximal output”. The default settings are
preset to 0% and 100%.