BioPAT
®
Trace | Multi Trace Operating Instructions 97
Operation
8.3.8.1 Events
In the submenu “Events”, individual actions are entered at certain times. For
every action (one shot measurement, calibration, start, standby) one or more
time points can be determined. These events are always handled in priority to
other actions (from the schedule or the parameter list). In the event that the
scheduled time of an action is conflicting with an action of the schedule or in
the parameter list, it will wait until this is finished.
It is possible to switch the device at a specific time to start (“Start”) and later
again in standby (“In standby”). During the measurements either the workflow
is executed (if enabled, see below “8.3.8.2 Workflow”, page 98) or the
measurement frequency and possibly edited autocalibrations according to
the parameter list.
If several test ports (BioPAT
®
Multi Trace) are present and enabled, the
respective measurements and the desired test port (1–4) can be assigned. The
number of individual measurements (corresponding to “Group” in the
parameter list) at the specified time is set via the input field “Number of
Measurements”.