BioPAT
®
Trace | Multi Trace Operating Instructions 99
Operation
The workflow is enabled by a check mark “Workflow active” in the selection
box. If this box is not checked, the measurement processes in the parameter
list are processed instead.
The selected action for the first time interval in line 1 is active only during the
second pass of the list. This ensures that the processing of the list begins right
after the start. On the second pass the selected first time interval in line 1
applies.
The number of individual measurements (corresponding to “Group” in the
parameter list) is set via the input field “Number of Measurements”. If several
test ports (BioPAT
®
Multi Trace) are present and enabled, the respective
measurements and the desired test port (1–4) can be assigned.
By pressing the “Attach” button the selected settings are transferred to the
right list box and entered in the bottom row. If an action has to be placed
between two existing entries, the lower line must be marked. Then, pressing
the “Paste” button inserts the selected setting. The entry can be marked by
clicking on a line and removed from the list again by pressing “Delete”.
Pressing the “Clear” button all entries will be deleted from the list box.
Pressing the “OK” button, all records are saved and activated. Press
“Canceled” to discard the changes.
The workflow is executed cyclically until the unit is stopped or put into
standby.