Choosing a Suitable Location
  In order to ensure that the analyzer 
functions properly, select a location that 
meets the following requirements:
§  Permissible ambient environment
  –   Temperature:   
+15 to +40°C (59°F to 104°F)
  –   Relative  humidity:   
10 – 85%, non-condensing
  –   Altitude:   
0 to 2,000 meters (0 to ~6562 ft)
§  The analyzer is intended for indoor use�
§  Place the analyzer on a rigid, horizontal 
surface, preferably free of vibration�
§  Avoid drafts and excessive temperature 
fluctuation�
§  Locate the analyzer away from  
flammable materials�
Connecting the Moisture Analyzer  
to AC Power
  This Class I Equipment (grounded  
type) analyzer contains a universal  
self-adjusting power supply rated to 
100-120/200-250V~5/2�5A 50/60 Hz� 
Check that this voltage matches your 
local line voltage� If your voltage is not 
within this rating, do not connect the 
power cord to the power supply�
Danger
!
 The analyzer may only be oper-
ated using the original power cord 
supplied� Please do not use it with 
an extension cord� The wall socket 
must be grounded� The AC Power 
mains supply fluctuations are not 
to exceed ±10% of the nominal 
supply voltage� The (Over Voltage) 
Category II for transient over volt-
ages is EN 61010-1� The pollution 
Degree 2 is supplied EN 61010-1� 
The analyzer is rated to Normal 
Protection (not protected against 
harmful ingress of moisture)�
§  Insert the power cord that with 
the analyzer into the power input 
receptacle on the back of the  
heater module�
§  Then plug the power cord into a 
wall socket�
§  Turn the analyzer on by moving 
the power switch on the back of 
the analyzer to the On position� 
The analyzer will briefly proceed 
through a self-diagnostic routine 
and then display the standby 
screen� It will automatically enter 
a warm-up period as noted on the 
screen� Should the analyzer display 
any warnings, consult the trouble-
shooting section of this manual,  
or call Sartorius�
13